Ground-based and Airborne Instrumentation for Astronomy XI
The core of HARMONI is the Integral Field Spectrograph (IFS), which consists of several subsystems including the IFS Pre-Optics (IPO). The IPO main objective is to receive the incoming light and reformat and condition it to provide a suitable input for the rest of the spectrograph. The IPO includes the Scale Changer Turret (SCT), a cryogenic mechanical turret that selects between different spatial scales or a pupil-imaging mode by inserting mirrors into the optical path or leaving it clear.
This paper reports on the testing campaign of the HARMONI IFS Pre-Optic's (IPO) Scale Changer Turret (SCT) prototype. The primary aim of these tests was to validate the proposed mechanical design by comparing alternative hardware and control configurations and confirm the overall system performance. We present a detailed analysis of the system and the laboratory setup.
The study compares the alternative software and hardware configurations to identify the best control-performance trade-offs. We discuss the results of iterative testing involving different motor types (standard versus cryogenic), various image acquisition configurations, and motor step configurations. The results were verified using the integration of the ELT ICS Function Control Framework (FCF), providing a realistic setup with the final software. This framework also ensures enhanced programming and testing efficiency. The tests demonstrate that the SCT design equipped with the HARMONI 1.5 hard-detent stop mechanism satisfies the positioning requirements under warm laboratory conditions, and they provide confidence for the forthcoming cryogenic validation.